1.2 Keys and Knobs
Vertical Axis and Channels
CH1 to CH4 Keys Section 5.1
Display menus for turning analog signal input channels on and off, for expanding and
reducing the vertical axis, and for setting the vertical position, coupling, probe type, offset
voltage, bandwidth limit, linear scaling, and waveform labels. Also, press any of these
keys to select which channel the SCALE and POSITION knobs will adjust. The channel
keys illuminate when their corresponding channel displays are on.
LOGIC Keys Section 5.2
Displays a menu for configuring the logic signal display (grouping, order, bus display, and
state display), skew adjustment, threshold levels, labels, etc. Press this key and then the
POSITION knob to set the vertical display position of the logic signal. Press this key and
then the SCALE knob to set the vertical display range of the logic signal.
POSITION Knob Sections 5.1 and 5.2
Before you turn this knob, select the channel that you want to adjust by pressing LOGIC
or a key from CH1 to CH4.
When a channel from CH1 to CH4 is selected, you can use this knob to adjust the
vertical display position (vertical position). When you change the vertical scale, the
location of the vertical position does not change. This knob has a push switch. You can
press the knob to reset the position to 0.0 divisions.
When the LOGIC channel is selected, you can adjust the vertical position of the LOGIC
channel.
SCALE Knob Sections 5.1 and 5.2
Use this knob to set the vertical scale. Before you turn this knob, select the channel that
you want to adjust by pressing LOGIC or a key from CH1 to CH4. If you change the
scale while waveform acquisition is stopped, the waveform is expanded or reduced
vertically. If you restart signal acquisition, the DL6000/DLM6000 acquires signals using
the new vertical scale.
This knob has a push switch. If you are adjusting a channel from CH1 to CH4, you can
press the knob to switch the resolution. Pressing the knob so that the Fine indicator
illuminates enables you to make finer adjustments.
If you are adjusting the LOGIC channel, you can switch between four different vertical
scales.
Horizontal Axis
POSITION Knob Section 6.2
You can change the horizontal display position (trigger position) of acquired waveforms.
When you change the time axis scale, the location of the trigger position does not
change. This knob has a push switch. You can press the knob to reset the position to the
default value of 50%.
DELAY Key Section 6.2
The DELAY key illuminates when you press it. While the key is illuminated, you can set
the trigger delay using the POSITION knob.
You can reset the trigger delay to its default value of 0 s by pressing the POSITION knob
when the DELAY key is illuminated.
TIME/DIV Knob Section 5.3
Use this knob to set the time scale. If you change the scale while waveform acquisition is
stopped, the scale change actually takes effect when you restart waveform acquisition.
HORIZONTAL
POSITION
TIME/DIV
PUSH
50%
0 s
DELAY
VERTICAL
POSITION
PUSH
ODIV
PUSH
FINE
SCALE
1
2
3
4
LOGIC