6-26 IM 701450-01E
Setting the Status of Each Channel
6. Press the Set Pattern soft key. The Set Pattern dialog box opens.
7. Use jog shuttle & SELECT to set the status of each channel to H, L, or X (IN,
OUT, or X if Window is ON).
8. Use jog shuttle & SELECT to set Condition to True or False.
9. Press ESC. The Set Pattern dialog box closes.
Setting the Level, Hysteresis, Trigger Coupling, and HF Rejection
10. Press the Level/Coupling soft key. The Level/Coupling dialog box opens.
11. Like in the case of the A->B(N) trigger, set the level, hysteresis, trigger coupling,
and HF rejection.
• The setup procedure is common to the A->B(N) trigger. See page 6-16.
• If Window is ON, set the window position and width, trigger coupling, and HF rejection.
The setting is the same as the window trigger. See section 6.13.
12. Press ESC. The Level/Coupling dialog box closes.
Note
The trigger level, hysteresis, trigger coupling, and HF rejection settings apply to both simple
and enhanced triggers.
Setting the Determination Time
13. Press the Time soft key.
If Width Type is T1<Pulse<T2, press the Time1/Time2 soft key to set the jog shuttle
control to Time1 or Time2.
6.11 Setting the Width (Pulse<Time, Pulse>Time, T1<Pulse<T2, and Time Out) Trigger (ENHANCED)