10-47
IM 701450-01E
10
Analyzing and Searching Waveforms
Menu on the DL7480Item Setup dialog box
Mark at the left of the highlighted
measurement parameter
Setting the Delay Measurement between Waveforms
9. Press the Delay Setup soft key. The Delay Setup dialog box opens.
Setting the Measurement Mode and Detection Conditions of the Waveform to Be
Measured
10. Use jog shuttle & SELECT to select the measurement mode (Mode box).
11. Press the CH1 to CH8/4, Math1, or Math2 soft key to select the waveform to be
measured.
• On the DL7440, select CH1 to CH4, Math1, or Math2.
• On the DL7480, select CH1 to CH8, Math1, or Math2. CH7, CH8, Math1, and Math2
appear when you press the To Page 2 soft key.
12. Use jog shuttle & SELECT to select which slope edge, rising or falling, of the
waveform being measured is to be detected (Polarity box under Measure).
13. Use jog shuttle & SELECT to set the edge detection count (the edge count to
which delay is measured, Edge Count box under Measure).
Setting the Reference Point
14. Use jog shuttle & SELECT to select the reference point (Reference box).
If you select Trig for the reference point, the succeeding steps are not necessary.
15. Use jog shuttle & SELECT to select the reference waveform (Trace box under
Reference).
16. Use jog shuttle & SELECT to select which slope edge, rising or falling, of the
reference waveform to be the reference (Polarity box under Reference).
17. Use jog shuttle & SELECT to set the edge detection count (the edge count to
be the reference point, Edge Count box under Reference) .
18. Press ESC. The Delay Setup dialog box closes.
Menu on the DL7480
Select the measurement mode.
Select which slope edge, rising or falling,
is to be detected.
Set the edge detection count
(the edge count to which delay is measured).
Select the reference point.
Select the reference waveform.
Select which slope edge, rising or falling,
is to be used as a reference.
Set the edge detection count
(which ed
g
e is used as the reference point).
10.6 Automated Measurement of Waveform Parameters

Table of Contents

Other manuals for YOKOGAWA DL7440

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL7440 and is the answer not in the manual?

YOKOGAWA DL7440 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL7440
CategoryTest Equipment
LanguageEnglish

Related product manuals