16-3
IM 701450-01E
16
Troubleshooting, Maintenance, and Inspection
Errors in Execution (600 to 799)
Code Message Corrective Action Sections
601 Invalid file name or SCSI ID. Check the file name or SCSI ID number. 12.4,
12.6 to 12.10
602, 603 No SCSI device or no storage media inserted. Check the SCSI device connection, SCSI ID number, 12.3, 12.4
and the presence of a storage medium.
604 Storage media failure. Check the storage medium.
605 File not found. Check the file name and storage medium Chapter 12
606 Storage media is protected. Turn OFF the write-protect switch on the storage
medium.
607 Storage media failure. Check the storage medium.
608-610 Invalid file name or SCSI ID. Check the file name or SCSI ID number. 12.4,
12.6 to 12.10
611, 612 Storage media full. Delete unneeded files or use another storage medium. 12.6,
12.14
613 Cannot delete a directory if there are files in Delete all the files in the directory to be deleted. 12.14
the directory.
614 File is protected. Change the attribute to R/W. 12.14
615 Physical format error. Reformat the storage medium. 12.6
If the same error occurs, the storage medium cannot
be formatted using the DL7400.
616-620 File system failure. Check using a different storage medium. If it still fails,
servicing required.
621 File is damaged. Check the file.
622-641 File system failure. Check using a different storage medium. If it still fails,
656-663 servicing required.
642 No storage media exists in SCSI device. Check the presence of the storage medium of the
SCSI device.
646-653 Storage media failure. Check the storage medium.
654 Storage media failure. Check the format type of the floppy disk. 12.6
665 Cannot load this file format. Files saved on other models (such as YOKOGAWA
DL/AG Series) cannot be loaded.
666 File is now being accessed. Wait a moment. Execute after the access is finished.
667 Cannot be executed while data acquisition is Press the START/STOP key to stop the waveform 7.1
in progress. acquisition first.
668 Cannot find “.HDR” file. Check the file. 12.8
669 Cannot load the specified file on this ROM Upgrade the firmware.
version or this model.
671 Save data not found. Check the existence of the data to be saved.
673 SCSI controller failure. Servicing required.
676 Unknown file format. Check whether the data is in a format that the 12.7 to
DL7400 can handle. 12.13
Change the extension
677 P-P compression cannot be used to save FFT Turn OFF the P-P compression and save the data. 12.8
waveforms.
679 Data that have been P-P compressed and
saved cannot be loaded.
680 Illegal printer head position. Move the release arm to the HOLD position. 11.1
681 Paper empty. Install the roll paper. 11.1
682 Printer overheated. Turn OFF the power immediately.
Servicing required.
683 Printer overheated. Turn OFF the power immediately.
Servicing required.
684 Printer is not installed. Check the specifications to see if your model has Page ii
the optional printer.
685 Printer time out. Servicing required.
16.2 Messages and Corrective Actions