16-4 IM 701450-01E
Code Message Corrective Action Sections
691 The printer has malfunctioned. Servicing required.
692 Printer error.
Turn the power of the printer from OFF to ON.
693 Printer offline.
694 Out of paper.
695 Printer is in use.
696 Cannot detect printer.
Turn ON the printer.
Check connectors.
697 No files supporting the thumbnail display window. 12.13
701 Cannot be executed while data acquisition is Press the START/STOP key to stop the waveform 7.1
acquisition first.
703 There is no data to be undone. Undo is not possible because the data immediately 4.4, 4.5
before the initialization or auto setup does not exist.
704 Cannot be executed while data acquisition is Press the START/STOP key to stop the waveform 7.1
acquisition first.
705 This data cannot be backed up.
706 There is no data to be recalled. 12.1
707 Cannot start during data output. Wait until the output is complete. Chapter 11
711 Cannot access file while hard-copying. Wait until the output is complete. 11.2
712 Cannot compress this screen image. Turn OFF the compression setting. 12.12
Turn off the compression switch.
713 Calibration failure. Servicing required.
Disconnect the input and execute again.
If it fails again, servicing is required.
726 Cannot start when loading waveform data that Execute unload. 12.8
has been saved in history All mode.
727 Insufficient output data. Increase Mag or widen the Time Range interval. 11.2
Increase Mag or widen the Time Range interval.
728 Hard copying. Press COPY again to abort. 11.2
Abort or wait until it is complete.
729 Cannot perform calibration while waveform data Unload the loaded waveform using the FILE menu. 12.8
is loaded.
730 Pattern is not specified. Set at least one of the search pattern to a value 10.4
other than ×.
731 Cannot start when waveform data that has been Unload the loaded waveform using the FILE menu. 12.8
acquired in the linear average mode is loaded.
732 Cannot be executed while computation is Turn OFF Display of the MATH menu and abort 9.1
in progress. the computation.
733 Failed to measure statistics. Check to see that waveform data to be measured 10.7
The target waveform data exists or the exists and that there is at least one cycle of waveform
measured waveform data may not exist. If Cycle within the measurement range.
Statistics is specified, the instrument may be
configured in a way that fails to detect the cycle.
734 Cannot be executed during the interleave mode. 10.11
Change the Clock CH.
735 Cannot store because the data is locked. Release the lock through Store Detail. 12.1
Release the lock through Store Detail.
736 The File item is inappropriate. Select Waveform, Snap, or Measure.
Select Waveform, Snap, or Measure.
737 Executing file Load, Save, or Format.
Abort or wait until it is complete.
738 Hard copying or saving image.
Abort or wait until it is complete.
16.2 Messages and Corrective Actions