CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 3.3 Platen Wall Adjustment
3 - 24
3
Search
Book TOC
Go Back
4. Ali
g
n Welded Platen:
a. Welded Platen walls cannot be strai
g
htened. If walls are not square, replace Platen.
b. It may be necessary to adjust Rack Guide and/ or Beam Block for Tool to fit properly.
To Adjust Rack Guide:
- Loosen (2) screws attachin
g
Rack Guide to Platen.
- Insert Rack Guide Ali
g
nment Tool, adjust Rack Guide, and ti
g
hten hardware.
To Adjust Beam Block:
-Go to “VP - 3.5 Rack Clearance Adjustment” (Pa
g
e 3-28).
5. Go to “VP - 3.4 Rack Clearance Check” (Pa
g
e 3-26).