3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 3.6 X-Axis Solenoid to Guide Rod Check
3 - 33
3
Search
Book TOC
Go Back
tinue with
Step 5b
.
b. Install Rack Guide Ali
nment Tool on ri
ht side of Platen.
c. Press (
←
X
→
) to move X-Axis Solenoids to ed
e of tool.
d. Check distance between Tool and solenoids usin
a Feeler Gau
e. Distances should be the
same (± .005 in., 0.127mm).
e. If Check is OK,
o to “VP - 3.8 Rack to Laser Shield Check” (Pa
e 3-39).
f. If Check Fails,
o to “VP - 3.7 X-Axis Solenoid to Guide Rod Adjustment” (Pa
e 3-36).