CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 3.6 X-Axis Solenoid to Guide Rod Check
3 - 33
3
Search
Book TOC
Go Back
tinue with
Step 5b
.
b. Install Rack Guide Ali
g
nment Tool on ri
g
ht side of Platen.
c. Press (
X
) to move X-Axis Solenoids to ed
g
e of tool.
d. Check distance between Tool and solenoids usin
g
a Feeler Gau
g
e. Distances should be the
same (± .005 in., 0.127mm).
e. If Check is OK,
g
o to “VP - 3.8 Rack to Laser Shield Check” (Pa
g
e 3-39).
f. If Check Fails,
g
o to “VP - 3.7 X-Axis Solenoid to Guide Rod Adjustment” (Pa
g
e 3-36).