CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 6.1 Mixing Head Position Check
3 - 80
3
Search
Book TOC
Go Back
2. Perform Mix Head Ali
g
nment Check as follows:
E and F Loader
a. Put Sample Loader in Manual Mode.
1.
Press and Hold
MAN
and
START
.
2.
Turn Power
ON
, hold keys down until BEEPING stops.
b. Move X-Axis Solenoids to Left End Stop.
c. Insert Rack Guide Ali
g
nment Tool in Ri
g
ht End of Platen with cutout facin
g
up and to the left
(Fi
g
ure 3-38).