CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 6.1 Mixing Head Position Check
3 - 83
3
Search
Book TOC
Go Back
h. Mixin
g
Head should fit completely into Tool
without shifting tool or Rack
. No Play between
Tool and Mixin
g
Head. If Check Fails,
g
o to “VP - 6.2 Mixin
g
Head Position Adjustment” (Pa
g
e
3-85).
G Loader [Europe]
a. Put Loader in Manual Mode.
1.
Press and Hold
MAN
and
START
.
2.
Turn Power
ON
, hold keys down until BEEPING stops.
b. Move X-Axis Solenoids to Left End Stop.
c. Insert Rack Guide Ali
g
nment Tool in Ri
g
ht End of Platen (Fi
g
ure 3-38).
d. Load Sarstedt Tube in Position # 4 of Rack and Place Rack on Left End of Platen (Fi
g
ure
3-40).
e. Press (
X
) to move Rack into cutout. Tube should be under Mixin
g
Paddle (Fi
g
ure 3-38).
f. Press (
Z
) to lower Mixin
g
Head onto Tube.
g
. Check Ali
g
nment of Tube.
1.
Press and hold (
R
) to Rotate Tube.
2.
Verify Tube is Centered in Rack while rotatin
g
(Fi
g
ure 3-40).
3.
If Check Fails,
g
o to “VP - 6.2 Mixin
g
Head Position Adjustment” (Pa
g
e 3-85).
4.
If Check is OK,
g
o to “VP - 6.3 Mixin
g
Home Sensor Check/Adjustment” (Pa
g
e 3-89).