28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 6.1 Mixing Head Position Check
3 - 83
3
Search
Book TOC
Go Back
h. Mixin
Head should fit completely into Tool
without shifting tool or Rack
. No Play between
Tool and Mixin
Head. If Check Fails,
o to “VP - 6.2 Mixin
Head Position Adjustment” (Pa
e
3-85).
G Loader [Europe]
a. Put Loader in Manual Mode.
1.
Press and Hold
MAN
and
START
.
2.
Turn Power
ON
, hold keys down until BEEPING stops.
b. Move X-Axis Solenoids to Left End Stop.
c. Insert Rack Guide Ali
nment Tool in Ri
ht End of Platen (Fi
ure 3-38).
d. Load Sarstedt Tube in Position # 4 of Rack and Place Rack on Left End of Platen (Fi
ure
3-40).
e. Press (
X
→
) to move Rack into cutout. Tube should be under Mixin
Paddle (Fi
ure 3-38).
f. Press (
Z
↓
) to lower Mixin
Head onto Tube.
. Check Ali
nment of Tube.
1.
Press and hold (
R
) to Rotate Tube.
2.
Verify Tube is Centered in Rack while rotatin
(Fi
ure 3-40).
3.
If Check Fails,
o to “VP - 6.2 Mixin
Head Position Adjustment” (Pa
e 3-85).
4.
If Check is OK,
o to “VP - 6.3 Mixin
Home Sensor Check/Adjustment” (Pa
e 3-89).