VERIFICATION PROCEDURES
VP-15.1 Barcode Reader Configuration (Europe Only)
3 - 164
3
Search
BookTOC
GoBack
7.Insert
Service Disk
into Data Station floppy disk drive.
8.Turn Sample Loader and Data Station power
ON
.
9.At
A:>
prompt, type
VBCDIAG
and press
[ENTER]
.
10.Data Station displays:
• Function Key Information
• RS-232 Configuration
CAUTION:If the message “PORT 1: UNABLE TO GO ONLINE ... COMM STATUS” displays,
repeat Step 1.
NOTE:If procedure fails, check cable wiring according to Figure 3-88, Page-3-167. If problem
persists, perform IP-17 and IP-21.
11.Change Baud Rate to 1200.
a.Press
[F3]
and wait 20 seconds.
b.Verify 1200 Baud Rate is displayed.
c.To keep existing Barcode settings, continue with Step 12.
d.To change Barcode settings to default values, press
[F3]
again.
System beeps.
12.Edit Barcode parameters.
NOTE:System responds slowly. Observe screen while entering commands and confirm
acceptance by system.