1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
Appendix 3 A - Sensors, LEDs and POTs
3 - 170
3
Search
Book TOC
Go Back
Mixer, Limit N/A CE MARK: CN 5, PIN
10
NON-CE MARK: TP
Header, Pin 4
Blocked: 5V
UnBlocked: 0V
N/A 6.5
Needle, Home N/A CE MARK: CN 5, PIN
3
NON-CE MARK: TP
Header, Pin 13
Blocked: 5V
UnBlocked: 0V
N/A 8.1
Needle, Limit N/A CE MARK: CN 5, PIN
9
NON-CE MARK: TP
Header, Pin 15
Blocked: 5V
UnBlocked: 0V
N/A 8.1
Tube Detect
(Beam Block)
Man See VP White Beam Block
UnBlocked: 1.0 -
1.3V
Black Beam Block
UnBlocked: 2.0V
Blocked: 0V
See
VP
5.1
5.2
X-Axis, Home N/A CE MARK: CN 4, PIN
3
NON-CE MARK: TP
Header, Pin 3
Blocked: 5V
UnBlocked: 0V
N/A 3.16
3.17
Table 3 - 7
Sensor LED Test Point Normal Voltages POT VP#