6241A/6242 DC Voltage Current Source/Monitor Operation Manual
5.4.2 Operational Principles
5-72
• Voltage range switching is done by A
5
, and the voltage measurement, voltage source, and the volt-
age limiter always take place in the same range.
•The A
5
and A
6
amps have high input impedance to minimize leakage.
•The A
7
amps also has a high input impedance to reduce error for 4-wire connection.
• The AD converter employs an integral type AD, and the integration time can be set between 100 μs