6241A/6242 DC Voltage Current Source/Monitor Operation Manual
4.2.6 MENU Key (Parameter Setting)
4-5
LMT Input Selects the limiter HL and LL value setting.
±Balance :
Both positive and negative values of HL and LL value
change at the same time.
Separate :
Sets HL and LL value separately.
The HL and LL set range is (HL value -(minus) LL
value) > (Minimum setting range).
The same polarities can be set.
Response Selects the source response.
Fast : Fast mode
Slow: Slow mode
SWEEP Sets the sweep source operations.
Sweep Type Selects sweep source mode type.
Linear: Executes a linear sweep.
Fixed: Executes a fixed sweep.
Random: Executes a random sweep.
Linear 2: Executes two slope linear sweep.
SWP Range Selects the range functions when sweeping.
Auto: Sweeps at each step from the start value to the stop
value, each in an optimum range.
Fix: Sweeps in the minimum fixed range, which can output
any source values, from the start value to the stop value.
Reverse Switches the reverse mode (double sweep) ON and OFF.
On: Sweeps from the start value to the top value and
continues sweeping back to the start value.
Off: Sweeps from the start value to the top value and then
stops.
Repeat Cnt Sets the number of times to repeat the sweep.
Setting 0 repeats indefinitely.
When a number between 1 to 1000 is set and the reverse mode is
on, each round sweep is counted as 1.
Rtrn Bias Selects whether the source value returns to the bias value or stays
at the existing value.
On: Returns to the bias value when the sweep is completed.
Off: Stays at the source value when the sweep is completed.
SWEEP VAL Sets values used when performing a sweep.
Parameters vary depending on the sweep type.