4 Modify the waveform.
Press [
Analys
][
DATA EDIT
][
EDIT D1
].
Press [
MODIFY START Y
][
1
][
ENTER
].
Press [
MODIFY STOP Y
][
1
][
ENTER
].
Press [
START X
][
1
][
0
][
mS
].
Press [
MODIFY START Y
][
0
][
ENTER
].
Press [
MODIFY STOP Y
][
0
][
ENTER
].
5 Configure the display.
Press [
Scale
], then press[
AUTOSCALE ON OFF
] to highlight ON.
6 Configure the arbitrary source.
Press [
Source
][
MORE CHOICES
][
ARB SRC SETUP
][
DATA REG D1
].
Press [
Rtn
].
Press [
ARBITRARY (D1 - D8)
].
The analyzer defaults to the nearest time record length, in this case 15.625 mS.
To build a different type of step signal, simply replace the start and stop values for the
X-andY-axisinstep6withyourown.
A: D1Time X:0 s Y:1 V
0s 15.6097ms
1.4
V
-600
mV
Real
200
mV
/div
Per Div: 200m Ref Pos: Cntr
Start X = 0
Start Y, Stop Y = 10
Start X = 10
StartY,StopY=0
Agilent 35670A
Operator's Guide The Arbitrary Source Option 1D4