Tosetupawaterfalldisplay
1 Activate the trace you want to use for a waterfall display.
2 Press [
Disp Format
][
WATERFALL SETUP
][
WATERFALL STEPS
] <number> [
ENTER
]to
specify the number of measurements you want to store in the waterfall buffer.
3 Press [
Z-AXIS RANGE
], then enter the range of measurements you want to observe in
the waterfall display.
4 Press [
WATERFALL ON OFF
] to highlight ON.
5 If the analyzer is not currently measuring, press [
Start
].
When you set [
WATERFALL ON OFF
] to ON, the analyzer adds a new trace to the upper
portion of a waterfall display each time a measurement (or an averaged set of
measurements) is completed. Older traces flow toward the bottom of the display
through its lower portion. The value you enter under [
Z-AXIS RANGE
] determines how
many traces will be displayed in the lower portion of the waterfall display.
The analyzer uses a waterfall buffer to store the results of the last N measurements (or
the last N averaged sets of measurements). You specify the value of N under
[
WATERFALL STEPS
]. After you pause a measurement, you can use the waterfall
markers—described later in this chapter—to go back and analyze the traces in the
waterfall buffer.
The Z-axis range
determines how many
traces appear in this
part of the
waterfall display.
Agilent 35670A
Operator's Guide Formatting the Display