To specify a weighting function
1 Press [
Analys
][
DATA EDIT
][
EDIT D7
].
2 Press [
START X
] <number> <unit>.
3 Press [
STOP X
] <number> <unit>.
4 Press [
MODIFY START Y
] <number> <unit>.
5 Press [
MODIFY STOP Y
] <number> <unit>.
The weighting function is stored in a dedicated data register. The default register is
D7. If you change the weighting register, your selection is stored in RAM. If you
preset the analyzer, the selection returns to the default data register, D7.
Caution The data registers are cleared when you turn off the analyzer. Save the data register to
afilebeforepower-downoritwillbelost.
The curve fitter does not modify the weight register when user weight is selected. For
more information, see “The Weighting Function,” presented earlier in this chapter.
Use [ DATA EDIT] to edit
aweightingfunction
Agilent 35607A
Curve Fit Option 1D3 Operator's Guide
16-32

Table of Contents

Other manuals for Agilent Technologies 35670A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 35670A and is the answer not in the manual?

Agilent Technologies 35670A Specifications

General IconGeneral
BrandAgilent Technologies
Model35670A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals