To measure performance parameters
1 Measure a step response.
2 Press [
Pause-Cont
].
3 Press [
Marker Fctn
][
TIME PARAMTERS
].
Press [
START TIME
] <number> <unit>.
or
Use the knob to move the marker to the appropriate start time.
4 Press [
STOP TIME
] <number> <unit>.
or
Use the knob to move the marker to the steady-state level.
5 Press [
OVERSHOOT
].
Press [
RISE TIME
].
Press [
SETTLING TIME
].
Press [
DELAY TIME
].
Agilent 35670A
Measuring Control Systems Operator's Guide