To transfer a curve fit table to a synthesis table
1 Press [
Analysis
][
SYNTHESIS
].
2 Press [
TABLE ON OFF
] to highlight ON.
3 Press [
COPY FROM CURVE FIT
].
The curve fit table transfers as a pole-zero format synthesis table. The analyzer’s
curve fitter does not work with a non-Hermitian frequency-responses because real,
measured systems are Hermitian-symmetric.
Any values already in the synthesis table are erased when the curve-fit table is copied.
Agilent 35607A
Curve Fit Option 1D3 Operator's Guide
16-38