To select the amplitude unit
1 Activate the trace whose amplitude unit you want to select.
2 Press [
Trace Coord
][
Y UNITS
], then press one of the following: [
VOLT
], [
VOLT^2
],
[
V/rtHz
], [
V^2/Hz (PSD)
], or [
V^2s/Hz (ESD)
].
3 Press [
AMPLITUDE PK PP RMS
] to highlight one of the following: PK to display peak
amplitude values, PP to display peak-to-peak values, or RMS to display
root-mean-square values.
If you have transducer units turned on, each occurrence of VOLT or V in the Y Units
menu is replaced by EU. EU is an abbreviation for engineering units—an alternate
name for transducer units. The next task tells you how to turn on transducer units.
Note: The options you select under the [
Meas Data
] and [
Trace Coord
] hardkeys can
limit the availability of Y-axis units.