EnteringaTimeDelay
When necessary, you can enter a time delay for a synthesis calculation. Time
delay can be used to simulate computational delay in a sampled system or
transport delay. Negative delays correspond to an advance.
The delay can be in terms of microseconds (uS), milliseconds (mS), or seconds
(S). The default is 0 seconds, which has no effect on the synthesis. If you
enter a non-zero time delay (t), the H(s) synthesis includes multiplication by:
Where:
f is frequency in Hertz
t is time delay
The time delay appears at the bottom of the table. A delay in the time domain
will appear as a linear phase ramp in the frequency domain plots.
e
j2 ft−π