EasyManua.ls Logo

Agilent Technologies 35670A

Agilent Technologies 35670A
322 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Connect the transducer’s output cable to the analyzer as described at the beginning of
this chapter, then secure the transducer to the calibrator.
4 Specify the measurement parameters.
Press [
Freq
][
SPAN
], then press the down-arrow key if you want to reduce the
frequency span.
Press [
Window
][
FLAT TOP
] to maximize amplitude accuracy.
5 Press [
Scale
], then press [
AUTOSCALE ON OFF
] to highlight ON.
6 Turn on the calibrator, then press [
Start
].
7 After the display has been updated at least once, move the marker to the calibrator’s
output frequency and enter calibrator’s output amplitude.
Press [
Marker
][
MARKER TO PEAK
].
Press [
Input
][
XDCR UNIT CHx SETUP
][
CAL VALUE AT MARKER
], then type the
calibrator’s specified output amplitude and press [
ENTER
].
The analyzer uses the number you enter in [
CAL VALUE AT MKR
] to calculate a new
value for [
XDCR SENSITVTY
]. This new value is the calibration factor that ensures the
accuracy of your transducer measurements. The analyzer automatically stores this
value and the [
XDCR UNIT LABEL
] value in nonvolatile memory, so they will not be
lost at preset or power-down. However, the values will be used only when
[
XDCR UNIT ON OFF
] is set to ON.
Y-axis readout matches
calibration amplitude after you
complete this task.
Agilent 35670A
Operator's Guide Measuring Rotating Machinery
1-5

Table of Contents

Other manuals for Agilent Technologies 35670A

Related product manuals