EasyManua.ls Logo

Agilent Technologies 8960 Series

Agilent Technologies 8960 Series
786 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
485
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_read.fm
READ
READ:ATXPower[:ALL]?
READ:CAPPower[:ALL]?
READ:CCTPhase[:ALL]?
Function Initiates and fetches an analog transmitter power measurement as a sequential operation.
Measurement results are returned in the following order: integrity, average analog power.
Query Integrity
Range: 0 to 23 see Integrity Indicator
Resolution: 1
Average analog transmitter power
Range: -100.00 to 100.00 dBm, 9.91 E+37
Resolution: 0.01 dB
Function Initiates and fetches an access probe power measurement as a sequential operation.
Measurement results are returned in the following order: integrity, access probe power.
Query Integrity
Range: 0 to 23 see Integrity Indicator
Resolution: 1
Access probe power
Range: -100.00 to 100.00 dBm, 9.91 E+37
Resolution: 0.01 dB
Function Initiates and fetches a code channel time and phase error measurement as a sequential
operation.
Measurement results are returned in the following order: integrity, time error pass/fail
indication, phase error pass/fail indication.
Query Integrity
Range: 0 to 23 see Integrity Indicator
Resolution: 1
Time error pass/fail indicator:
Range: 0 (pass), 1 (fail), 9.91 E+37
Phase error pass/fail indicator:
Range: 0 (pass), 1 (fail), 9.91 E+37

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Related product manuals