READ
READ:ATXPower[:ALL]?
READ:CAPPower[:ALL]?
READ:CCTPhase[:ALL]?
Function Initiates and fetches an analog transmitter power measurement as a sequential operation.
Measurement results are returned in the following order: integrity, average analog power.
Query Integrity
Range: 0 to 23 see Integrity Indicator
Resolution: 1
Average analog transmitter power
Range: -100.00 to 100.00 dBm, 9.91 E+37
Resolution: 0.01 dB
Function Initiates and fetches an access probe power measurement as a sequential operation.
Measurement results are returned in the following order: integrity, access probe power.
Query Integrity
Range: 0 to 23 see Integrity Indicator
Resolution: 1
Access probe power
Range: -100.00 to 100.00 dBm, 9.91 E+37
Resolution: 0.01 dB
Function Initiates and fetches a code channel time and phase error measurement as a sequential
operation.
Measurement results are returned in the following order: integrity, time error pass/fail
indication, phase error pass/fail indication.
Query Integrity
Range: 0 to 23 see Integrity Indicator
Resolution: 1
Time error pass/fail indicator:
Range: 0 (pass), 1 (fail), 9.91 E+37
Phase error pass/fail indicator:
Range: 0 (pass), 1 (fail), 9.91 E+37

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals