EasyManuals Logo

Agilent Technologies 8960 Series Reference Guide

Agilent Technologies 8960 Series
786 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #484 background imageLoading...
Page #484 background image
484
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_read.fm
READ
Diagram Conventions on page 248
READ:AFANalyzer[:ALL]?
Function Initiates and fetches an audio frequency analyzer measurement as a sequential operation.
Measurement results are returned in the following order: integrity, average audio level, average
SINAD, average audio distortion.
Query Integrity
Range: 0 to 23 see Integrity Indicator
Resolution: 1
Average audio level
Range: 0 to 20 volts, 9.91 E+37
Resolution: the lesser of 4 significant digits or 10 uV.
Average SINAD
Range: -99.00 to 99.00 dB, 9.91 E+37
Resolution: 0.01 dB
Average audio distortion
Range: 0.00 to 99.00%, 9.91 E+37
Resolution: 0.01%
READ
:WQUality
[:ALL]
?
(returns Integrity, Rho, Frequency Error,
Time Error, Carrier Feedthrough, Phase
Error, Magnitude Error, EVM)
:TFERror
[:ALL]
?
(returns Integrity, Frame Error Rate, Received Good
Frames, Received Blank Frames, Transmitted Good
Frames, Transmitted Blank Frames, Frame Error
Rate Confidence)

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals