484
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_read.fm
READ
Diagram Conventions on page 248
READ:AFANalyzer[:ALL]?
Function Initiates and fetches an audio frequency analyzer measurement as a sequential operation.
Measurement results are returned in the following order: integrity, average audio level, average
SINAD, average audio distortion.
Query Integrity
Range: 0 to 23 see Integrity Indicator
Resolution: 1
Average audio level
Range: 0 to 20 volts, 9.91 E+37
Resolution: the lesser of 4 significant digits or 10 uV.
Average SINAD
Range: -99.00 to 99.00 dB, 9.91 E+37
Resolution: 0.01 dB
Average audio distortion
Range: 0.00 to 99.00%, 9.91 E+37
Resolution: 0.01%
READ
:WQUality
[:ALL]
?
(returns Integrity, Rho, Frequency Error,
Time Error, Carrier Feedthrough, Phase
Error, Magnitude Error, EVM)
:TFERror
[:ALL]
?
(returns Integrity, Frame Error Rate, Received Good
Frames, Received Blank Frames, Transmitted Good
Frames, Transmitted Blank Frames, Frame Error
Rate Confidence)

Table of Contents