491
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_read.fm
READ
READ:SAUDio[:ALL]?
READ:TFERror[:ALL]?
Function Initiates and fetches a swept audio measurement as a sequential operation.
Measurement results are returned in the following order: integrity, and a comma separated list
of average audio level and average audio distortion for each frequency point in the sweep.
For frequencies about 10 kHz, the measurement will return 9.91 E+37.
Query Integrity
Range: 0 to 23 see Integrity Indicator
Resolution: 1
Average Audio Level
Range: 1 mV to 20 V, 9.91 E+37
Resolution: the lesser of 4 significant digits or 10 uV
Average Audio Distortion
Range: 0 to 99.9%, 9.91E+37
Resolution: 0.1
Function Initiates and fetches the TDSO FER (test data service option (SO32) frame error rate)
measurements as a sequential operation.
FER measurement results are returned in the following order: integrity, frame error rate (%),
received good frames, received blank frames, transmitted good frames, transmitted blank
frames, frame error rate confidence (%).

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals