EasyManua.ls Logo

Agilent Technologies 8960 Series

Agilent Technologies 8960 Series
786 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
488
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_read.fm
READ
READ:DAPower[:ALL]?
READ:FM[:ALL]?
Function Initiates and fetches a digital average power measurement as a sequential operation.
Measurement results are returned in the following order: integrity, digital average transmit
power.
If the multi-measurement state is on, the result represents the average of a number of digital
average power measurements. Multi-measurement state, by default, is off. To set up
multi-measurements, see SETup:DAPower:COUNt[:SNUMber] .
Query Integrity
Range: 0 to 23 see Integrity Indicator
Resolution: 1
Digital average power:
Range: -100.00 to 100.00 dBm, 9.91 E+37
Resolution: 0.01 dB
Function Initiates and fetches an FM measurement as a sequential operation.
Measurement results are returned in the following order: integrity, average frequency deviation,
average distortion.
The distortion measurement must be enabled before a measurement result (other than
9.91E+37) is reported. See SETup:AFANalyzer:SDIStortion:STATe on page 510.
Query Integrity
Range: 0 to 23 see Integrity Indicator
Resolution: 1
Average FM frequency deviation
Range: 0.000 to 99.000 kHz, 9.91 E+37
Resolution: 0.001 Hz
Average FM distortion
Range :0.0 to 100.0%, 9.91 E+37
Resolution: 0.1%
Average FM frequency
Range: 0.0 to 99999.0 Hz, 9.91 E+37
Resolution: 0.1 Hz

Table of Contents