EasyManuals Logo

Agilent Technologies 8960 Series Reference Guide

Agilent Technologies 8960 Series
786 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #541 background imageLoading...
Page #541 background image
541
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_setup_fm.fm
SETup:FM
Diagram Conventions on page 248
SETup:FM:CONTinuous
Function Sets/queries the trigger arm state for the FM measurement. Determines whether measurement
will make one measurement then stop (0|OFF (single)), or automatically rearm upon completion
of one measurement and repeat the process (1|ON (continuous)).
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting 0 (off)
Programming Example
OUTPUT 714;"SETUP:FM:CONTINUOUS ON
!Sets measurement to continuous trigger mode.
SETup :FM :FILTer <sp><num value>[HZ|
KHZ|MHZ|GHZ
[:TYPE]
(returns NONE|TBP|CMES|BPAS50|
[:FREQuency]
<sp>NONE|TBPass|CMESsage|BPASs50|
:TBPass
? (returns num value)
BPASs300
BPAS300)
?
:FREQuency
:STATe
<sp>1|ON|0|OFF
? (returns1|0)
SETup :FM
:TIMeout
[:STIMe]
? (returns num value)
:TIME
? (returns num value)
:STATe
<sp>1|ON|0|OFF
Complex Command
? (returns1|0)
<sp><num value>[S|MS|US|NS]
<sp><num value>[S|MS|US|NS]

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals