589
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_status_operation.fm
STATus:OPERation:
STATus:OPERation:NMRReady:CDMA Register Bit Assignments
The STATus:OPERation:NMRReady:CDMA register bits indicate when a measurement has been completed
and new measurement results are available.
Bit
Number
Binary
Weighting
Condition Description
15 32768 Not Used. Defined by SCPI. This bit is always 0.
14 16384 Reserved for future use. This bit is always 0.
13 8192 Reserved for future use. This bit is always 0.
12 4096 Reserved for future use. This bit is always 0.
11 2048 Reserved for future use. This bit is always 0.
10 1024 Reserved for future use. This bit is always 0.
9 512 Reserved for future use. This bit is always 0.
8 256 CTXSpurious New Measurement
Result Ready
This bit is a 1 if the measurement has been
completed and has produced new results.
This bit is a zero at power on, after a preset and
while a measurement is in Measuring States. See
Triggering of Measurements on page 218.
7 128 GPOWer New Measurement
Result Ready
This bit is a 1 if the measurement has been
completed and has produced new results.
This bit is a zero at power on, after a preset and
while a measurement is in Measuring States. See
Triggering of Measurements on page 218.
6 64 TROPower New Measurement
Result Ready
This bit is a 1 if the measurement has been
completed and has produced new results.
This bit is a zero at power on, after a preset and
while a measurement is in Measuring States. See
Triggering of Measurements on page 218.
5 32 CAPPower New Measurement
Result Ready
This bit is a 1 if the measurement has been
completed and has produced new results.
This bit is a zero at power on, after a preset and
while a measurement is in Measuring States. See
Triggering of Measurements on page 218.
4 16 CFERror New Measurement
Result Ready
This bit is a 1 if the measurement has been
completed and has produced new results.
This bit is a zero at power on, after a preset and
while a measurement is in Measuring States. See
Triggering of Measurements on page 218.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals