STATus:QUEStionable
STATus:QUEStionable:CALL:TA2000 Condition Register Bit Assignment
The STATus:QUEStionable:CALL:TA2000 registers will contain information about which event(s) occurred
during call processing that indicate what call processing procedure failed
Program Example - STATus:QUEStionable:CALL:TA2000 Register Bit Assignment
OUTPUT 714;STATUS:QUESTIONABLE:CALL:TA2000:EVENT?
!Queries and clears the Questionable Call Event Register
OUTPUT 714;STATUS:QUESTIONABLE:CALL:TA2000:CONDITION?
!Queries and clears the Questionable Call Condition Register
OUTPUT 714;STATUS:QUESTIONABLE:CALL:TA2000:ENABLE 1024
!Sets the Questionable Call Enable Register for bit 10
OUTPUT 714;STATUS:QUESTIONABLE:CALL:TA2000:NTRANSITION 2
!Sets the Questionable Call Negative Transition Filter Register for bit 1
OUTPUT 714;STATUS:QUESTIONABLE:CALL:TA2000:PTRANSITION 2
!Sets the Questionable Call Positive Transition Filter Register for bit 1
Bit
Number
Binary
Weighting
Condition Description
15 32768 Not Used. Defined by SCPI. This bit is always 0.
14 16384 MUI Maskable Message.
13 8192 Reserved for future use. This bit is always 0.
12 4096 Reserved for future use. This bit is always 0.
11 2048 Reserved for future use. This bit is always 0.
10 1024 Reserved for future use. This bit is always 0.
9 512 Reserved for future use. This bit is always 0.
8 256 Reserved for future use. This bit is always 0.
7 128 Reserved for future use. This bit is always 0.
6 64 Reserved for future use. This bit is always 0.
5 32 Reserved for future use. This bit is always 0.
4 16 MS reject order received This bit is a 1 if the Service Option or Radio
Configuration change was rejected by the MS while
connected.
3 8 Carrier not detected on new
channel
This bit is a 1 when no power is detected by the test
set after a handoff or when making a call.
2 4 Handoff completion not received This bit is a 1 if the test set does not receive a
handoff completion message from the MS.
1 2 Release order not received This bit is a 1 if the MS does not send the call release
to the test set.
0 1 Extension Bit This bit is always 0.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals