736
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_ts_err_msg_pos100.fm
+100 through +199 Error Message Descriptions
+100 through +199 Error Message Descriptions
Description
A device-specific error indicates that the instrument has detected an error that occurred because some
operations did not properly complete, possibly due to an abnormal hardware or firmware condition. For
example, an attempt by the user to set an out of range value will generate a device specific error.
These are general errors generated by the core instrument. When one of these errors is generated, the ‘+100
errors’ bit in the questionable error status register is set.
Error Message Description
+101 Assert; Cycle power. Assert message<message1> <message1> will appear as:
If the DSP generated the assert:
;P:DSP T:<task ID> E:<error code> C:<error
classif.> F1:<flag 1> F2:<flag 2>
If the Protocol processor generated the assert:
;P:Protocol T:<task ID> L:<line number> F:<file
name>
If the Host processor generated the assert:
;P:Host T:<task ID> L:<line number> F:<file
name>
+102 Exception; Cycle power. Exception
message<message2>
<message2> will appear as: T:<task ID> or V:<vector
number> or PC:<program counter> or DA:<data adrs
reg value>.
Vector number, program counter and data address
register values are hexadecimal format.
+103 Failure; No measurements or settings can be made Indicates none of the VI’s are operational because a
serious problem exists.
+104 Failure; No settling operations will take place Indicates none of the VI’s are operational because a
serious problem exists.
+105 Failure; No measurements or setting can be made
for the function selected
Indicates none of the VI’s are operational because a
serious problem exists.
+110 Input pacing; Internal communication queue
overflow likely
Indicates that GPIB commands are too fast for the
device input queue and should be slowed.
+111 Input pacing; Internal communication queue
overflow imminent. Pacing increased
Indicates that GPIB commands were too fast and the
device input queue has not kept pace.
+112 Internal error; Protocol error <message3> <message3> is an eight digit hexadecimal number that
is the error code reported by protocol.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals