Chapter 4 187
Calibration
Changing the Calibration Kit Definition
4. Calibration
Step 10. Press Label and input a new label for the standard using the keypad displayed on the
screen.
Step 11. Press Return.
Step 12. Repeat Step 6 to Step 11 to redefine all standards for which changes are necessary.
Step 13. Press Return.
Step 14. Press Specify CLSs.
Step 15. Select the class to be redefined according to the list below.
Softkey Function
Sub Class Selects a subclass you want to use.
Open Selects the OPEN class
Short Selects the SHORT class
Load Selects the LOAD class
Thru Selects the THRU class
TRL Reflect Selects TRL Reflect.
TRL Line/Match Selcts TRL Line/Match.
Step 16. For TRL Thru, select ports.
Softkey Function
Port 1-2 Selects port 1 and port 2.
Port 1-3 Selects port 1 and port 3.
Port 1-4 Selects port 1 and port 4.
Port 2-3 Selects port 2 and port 3.
Port 2-4 Selects port 2 and port 4.
Port 3-4 Selects port 3 and port 4.
NOTE Select Set All to use the same standards for all test ports.
Step 17. Select the test port according to the list below.
Softkey Function
Port 1 Selects port 1
Port 2 Selects port 2
Port 3 Selects port 3
Port 4 Selects port 4