196 Chapter 4
Calibration
Changing the Calibration Kit Definition
85032B
1. Short(m) 2. Open(m) 3. Broadband 4. Thru
Label Short(m) Open(m) Broadband Thru
STD Type Short Open Load Delay/Thru
C0 [×10
-15
F]
0 119.09 0 0
C1 [×10
-27
F/Hz]
0 -36.955 0 0
C2 [×10
-36
F/Hz
2
]
0 26.258 0 0
C3 [×10
-45
F/Hz
3
]
0 5.5136 0 0
L0 [×10
-12
H]
0 0 0 0
L1 [×10
-24
H/Hz]
0 0 0 0
L2 [×10
-33
H/Hz
2
]
0 0 0 0
L3 [×10
-42
H/Hz
3
]
0 0 0 0
Offset Delay [s] 93 f 0 0 0
Offset Z0 [Ω] 49.992 50 50 50
Offset Loss [Ω/s] 700 M 700 M 700 M 700 M
Arb. Impedance [Ω] 50 50 50 50
7. Short(f) 8. Open(f)
Label Short(f) Open(f)
STD Type Short Open
C0 [×10
-15
F]
0 62.14
C1 [×10
-27
F/Hz]
0 -143.07
C2 [×10
-36
F/Hz
2
]
0 82.92
C3 [×10
-45
F/Hz
3
]
0 0.76
L0 [×10
-12
H]
0 0
L1 [×10
-24
H/Hz]
0 0
L2 [×10
-33
H/Hz
2
]
0 0
L3 [×10
-42
H/Hz
3
]
0 0
Offset Delay [s] 17.817 p 17.411 p
Offset Z0 [Ω] 50.209 50
Offset Loss [Ω/s] 2.1002 G 700 M
Arb. Impedance [Ω] 50 50
NOTE The designations (m) and (f) in the name (label) of the standard indicate male (m) and
female (f) analyzer connectors.