Chapter 4 195
Calibration
Changing the Calibration Kit Definition
4. Calibration
85032F
1. Short(m) 2. Open(m) 7. Short(f) 8. Open(f)
Label Short(m) Open(m) Short(f) Open(f)
STD Type Short Open Short Open
C0 [×10
-15
F]
0 89.939 0 89.939
C1 [×10
-27
F/Hz]
0 2536.8 0 2536.8
C2 [×10
-36
F/Hz
2
]
0 -264.99 0 -264.99
C3 [×10
-45
F/Hz
3
]
0 13.4 0 13.4
L0 [×10
-12
H]
3.3998 0 3.3998 0
L1 [×10
-24
H/Hz]
-496.4808 0 -496.4808 0
L2 [×10
-33
H/Hz
2
]
34.8314 0 34.8314 0
L3 [×10
-42
H/Hz
3
]
-0.7847 0 -0.7847 0
Offset Delay [s] 45.955 p 41.19 p 45.955 p 40.8688 p
Offset Z0 [Ω] 49.99 50 49.99 50
Offset Loss [Ω/s] 1.087 G 930 M 1.087 G 930 M
Arb. Impedance [Ω] 50 50 50 50
9. Broadband 11. Thru 13. (f/f)Adapter 14. (m/m)Adapter
Label Broadband Thru (f/f)Adapter (m/m)Adapter
STD Type Load Delay/Thru Delay/Thru Delay/Thru
C0 [×10
-15
F]
0 0 0 0
C1 [×10
-27
F/Hz]
0 0 0 0
C2 [×10
-36
F/Hz
2
]
0 0 0 0
C3 [×10
-45
F/Hz
3
]
0 0 0 0
L0 [×10
-12
H]
0 0 0 0
L1 [×10
-24
H/Hz]
0 0 0 0
L2 [×10
-33
H/Hz
2
]
0 0 0 0
L3 [×10
-42
H/Hz
3
]
0 0 0 0
Offset Delay [s] 0 0 339 p 339 p
Offset Z0 [Ω] 50 50 50 50
Offset Loss [Ω/s] 0 0 2.2 G 2.2 G
Arb. Impedance [Ω] 50 50 50 50
NOTE The designations (m) and (f) in the name (label) of the standard indicate male (m) and
female (f) analyzer connectors.