Appendix D 665
Softkey Functions
Calibration Menu
D. Softkey Functions
(Continued)
Calibrate
Enhanced
Response
Select Port
3-1 (S31
S11)
*1
Selects the transmission test (measurement of S
31
) for test port
13.
:SENS{1-16}:CORR:COLL:METH:ERES 3,1
4-1 (S41
S11)
*2
Selects the transmission test (measurement of S
41
) for test port
14.
:SENS{1-16}:CORR:COLL:METH:ERES 4,1
1-2 (S12 S22) Selects the transmission test (measurement of S
12
) for test port
21.
:SENS{1-16}:CORR:COLL:METH:ERES 1,2
3-2 (S32
S22)
*1
Selects the transmission test (measurement of S
32
) for test port
23.
:SENS{1-16}:CORR:COLL:METH:ERES 3,2
4-2 (S42
S22)
*2
Selects the transmission test (measurement of S
42
) for test port
24.
:SENS{1-16}:CORR:COLL:METH:ERES 4,2
1-3 (S13
S33)
*1
Selects the transmission test (measurement of S
13
) for test port
31.
:SENS{1-16}:CORR:COLL:METH:ERES 1,3
2-3 (S23
S33)
*1
Selects the transmission test (measurement of S
23
) for test port
32.
:SENS{1-16}:CORR:COLL:METH:ERES 2,3
4-3 (S43
S33)
*2
Selects the transmission test (measurement of S
43
) for test port
34.
:SENS{1-16}:CORR:COLL:METH:ERES 4,3
1-4 (S14
S44)
*2
Selects the transmission test (measurement of S
14
) for test port
41.
:SENS{1-16}:CORR:COLL:METH:ERES 1,4
2-4 (S24
S44)
*2
Selects the transmission test (measurement of S
24
) for test port
42.
:SENS{1-16}:CORR:COLL:METH:ERES 2,4
3-4 (S34
S44)
*2
Selects the transmission test (measurement of S
34
) for test port
43.
:SENS{1-16}:CORR:COLL:METH:ERES 3,4
Cancel Returns to the softkey display screen one level higher.
Open Executes an OPEN calibration on the test port selected in the Select
Port
menu.
:SENS{1-16}:CORR:COLL:OPEN
Short Executes a SHORT calibration on the test port selected in the Select
Port
menu.
:SENS{1-16}:CORR:COLL:SHOR
Load Executes a LOAD calibration on the test port selected in the Select
Port
menu.
:SENS{1-16}:CORR:COLL:LOAD
Thru Executes a THRU response calibration on the test ports selected in
the
Select Ports menu.
:SENS{1-16}:CORR:COLL:THRU
Isolation
(Optional)
Executes an isolation calibration on the test ports selected in the
Select Ports menu.
:SENS{1-16}:CORR:COLL:ISOL
Done Terminates a calibration process and calculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:SAVE
Cancel Displays softkeys for canceling a calibration.
OK Cancels a calibration process and clears the measurement data of
standard.
:SENS{1-16}:CORR:COLL:CLE
Cancel Returns to the softkey display screen one level higher.
Return Returns to the softkey display screen one level higher.
Key Operation Function SCPI Command