Appendix D 671
Softkey Functions
Calibration Menu
D. Softkey Functions
(Continued)
Calibrate
3-Port TRL Cal
*1
Reflect Displays softkeys for executing reflection calibrations. Symbols x, y and z refer to the three test ports selected in
the
Select Ports menu (x, y, and z are always 1, 2, and 3, respectively, for models with Option 313 or 314).
Port x
Reflect
Executes a OPEN/SHORT reflection calibration on test port x.
:SENS{1-16}:CORR:COLL:TRLR x
Port y
Reflect
Executes a OPEN/SHORT reflection calibration on test port y.
:SENS{1-16}:CORR:COLL:TRLR y
Port z
Reflect
Executes a OPEN/SHORT reflection calibration on test port z.
:SENS{1-16}:CORR:COLL:TRLR z
Return Returns to the softkey display screen one level higher.
Line/Match Displays softkeys for executing LINE/MATCH calibrations. Symbols x, y, and z refer to the three test ports
selected in the
Select Ports menu (x, y, and z are always 1, 2, and 3 respectively, for models with Option 213 or
214).
x-y
Line/Match
Executes a LINE/MATCH calibration on test ports x and y.
:SENS{1-16}:CORR:COLL:TRLL x,y
:SENS{1-16}:CORR:COLL:TRLL y,x
x-y Fwd(Syx) Executes a forward LINE/MATCH calibration on test ports x and y.
:SENS{1-16}:CORR:COLL:TRLL y,x
x-y Rvs(Sxy) Executes a reverse LINE/MATCH calibration on test ports x and y.
:SENS{1-16}:CORR:COLL:TRLL x,y
x-z
Line/Match
Executes a LINE/MATCH calibration on test ports x and z.
:SENS{1-16}:CORR:COLL:TRLL x,z
:SENS{1-16}:CORR:COLL:TRLL z,x
x-z Fwd(Szx) Executes a forward LINE/MATCH calibration on test ports x and z.
:SENS{1-16}:CORR:COLL:TRLL z,x
x-z Rvs(Sxz) Executes a reverse LINE/MATCH calibration on test ports x and z.
:SENS{1-16}:CORR:COLL:TRLL x,z
y-z
Line/Match
Executes a LINE/MATCH calibration on test ports y and z.
:SENS{1-16}:CORR:COLL:TRLL y,z
:SENS{1-16}:CORR:COLL:TRLL z,y
y-z Fwd(Szy) Executes a forward LINE/MATCH calibration on test ports y and z.
:SENS{1-16}:CORR:COLL:TRLL z,y
y-z Rvs(Syz) Executes a reverse LINE/MATCH calibration on test ports y and z.
:SENS{1-16}:CORR:COLL:TRLL y,z
Return Returns to the softkey display screen one level higher.
Done Terminates a calibration process and calculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:SAVE
Cancel Displays softkeys for canceling a calibration.
OK Cancels a calibration process and clears the measurement data of
standard.
:SENS{1-16}:CORR:COLL:CLE
Cancel Returns to the softkey display screen one level higher.
Return Returns to the softkey display screen one level higher.
4-Port TRL Cal
*2
Displays softkeys for executing 4-port TRL calibrations.
Thru/Line Displays softkeys for executing THRU/LINE calibrations.
1-2 Thru/Line Executes a THRU/LINE calibration for test port 12.
:SENS{1-16}:CORR:COLL:TRLT 1,2
:SENS{1-16}:CORR:COLL:TRLT 2,1
1-3 Thru/Line Executes a THRU/LINE calibration for test port 13.
:SENS{1-16}:CORR:COLL:TRLT 1,3
:SENS{1-16}:CORR:COLL:TRLT 3,1
1-4 Thru/Line Executes a THRU/LINE calibration for test port 14.
:SENS{1-16}:CORR:COLL:TRLT 1,4
:SENS{1-16}:CORR:COLL:TRLT 4,1
2-3 Thru/Line Executes a THRU/LINE calibration for test port 23.
:SENS{1-16}:CORR:COLL:TRLT 2,3
:SENS{1-16}:CORR:COLL:TRLT 3,2
2-4 Thru/Line Executes a THRU/LINE calibration for test port 24.
:SENS{1-16}:CORR:COLL:TRLT 2,4
:SENS{1-16}:CORR:COLL:TRLT 4,2
3-4 Thru/Line Executes a THRU/LINE calibration for test port 34.
:SENS{1-16}:CORR:COLL:TRLT 3,4
:SENS{1-16}:CORR:COLL:TRLT 4,3
Return Returns to the softkey display screen one level higher.
Key Operation Function SCPI Command