Quick Start Guide 2
Agilent Nano Indenter G200 User’s Guide 2-24
Figure 2-30Microscope focus and brightness controls
Position Sample Under the Microscope
1 In handset area in the left panel, right-click and select Nano
HandSet from the resulting pop-up menu.
2 Click on the top left circle in the schematic map, which corresponds
to the rear left sample in the sample tray.
3 Right-click anywhere on the schematic map and select Move to
Target from the resulting pop-up menu.
4 Visually confirm that the microscope is over the rear left sample disc
in the sample tray. The microscope does not have to be directly over
the sample at this point.
5 Adjust the sample position until the light from the microscope shines
directly on the sample.
Locate the height-adjustment thumbwheel for the rear left sample disk
in the sample tray.
Figure 2-31 on page 2-25 shows the thumbwheel,
which is not visible from the front of the instrument. To access the
thumbwheel, you will have to reach behind the sample tray.
Do not adjust the focus controls in NanoSuite from this point forward
until the samples in the four corners of the sample tray have been set to
their proper heights, described next. Doing so could damage the
indenter tip.