141
Appendix
F
F
Option Value Description Save last
value
LogOffTimeout 20 Number of minutes of inactivity until the user is
automatically logged out. Applies only when AutoLogOff
is active.
MaxCellDensity 10000000 Maximum cell concentration (in cells/mL) per image.
When this limit is exceeded, the user will be requested to
dilute the sample.
This warning will only directly appear as a warning
box if the “ErrorLevelDisplayToUser” is set to “4”.
MaxSedimentation
Delay
2 Maximum acceptable delay (in seconds) between the
end of the sedimentation phase and the beginning of
the main scan. The user is warned if the time interval is
exceeded.
Measurement
CriticalTime
90 Maximum acceptable time delay during a critical
measurement phase (i.e., between the beginning
of sample preparation to the capturing of the last
image) when the measurement is interrupted with the
Emergency Stop. If this time is exceeded, the results are
not reliable (particularly the viability value) and the user
is warned.
MultiRunClear
ListDateTime
Date/time when the Multi Run list was cleared. ×
MultiRunClear
ResultListDateTime
Date/time when the Multi Run overall result list was
cleared.
×
OverallTypeFor
Exponential
Regression
ViableCell
Density
(not adjustable) Overall Type for the exponential
regression used for the determination of doubling time
and specific growth rate.
SamplePortName Auto Sampler System default sample port:
Auto Sampler/Single Port/Remote Port.
SampleTraySize 20 Number of Cedex Sample Cups that the Multi Sample
Tray can hold.
SchedulerTimer
Period
5 Time interval (in seconds) in which the Scheduler should
check its task list.
SSTFailureReaction Restrict Reaction to a failed checkpoint. Possible values include
Audit: only an entry in the Audit Trail.
Warning: an additional warning before running a
measurement.
Restrict: the system is blocked.
SSTTimerPeriod 300 Time interval (in seconds) in which the SST checkpoints
are checked for validity.
SSTTrending
MovingAverage
3 Number of results used for the calculation of the
“moving average” for the SST Trending Chart.
SSTWorkarea Standard Default Workarea where the SST measurements are
stored.
StreamLiveImages 0 Use of “live image” during cleaning routines in the
Cedex Standard System (1=yes, 0=no).
TrayTypeNumber 0 Sample Tray type used for the Multi Sampler.
System Options and User Preferences