142
Cedex HiRes Analyzer Operator’s Manual
F
F
Settings in the “Measurement” tab:
Option Value Description Save last
value
AskBeforeMultiRun 0 Require to confirm the start of a multi run.
CellDiameter
HistogramDividers
(0) (0) (0) Divider settings for use in the cell diameter histogram. ×
CellType Std. Size Default Cell Type for measurements. ×
CellTypeFocus
Offset
-25 Default focus offset for the creation of new Cell Types if
the new Cell Type is not based on a previous Cell Type.
CellType
Sedimentation
Duration
60 Default sedimentation time (in seconds) for the creation
of new Cell Types if the new Cell Type is not based on a
previous Cell Type.
DefaultCombined
Number
1 Number of images that should be added together
for a summary of the measurements results in the
“Measurement Statistics” Histogram.
Mask image
required
0 Mask image required.
Measurement
Precision
Normal Default precision setting for the measurement. ×
Measurement
PrescanRequired
1 Prescan required for a Measurement (0-False/1-True).
ObjectDiameter
HistogramDividers
(0) (0) (0) Divider settings for use in the Object Diameter
Histogram.
×
PercentageOf
ValidImages
75.0 Percentage value of images which must be valid to
ensure the measurement is valid.
×
ReactorId Default ReactorID for a new measurement. ×
SampleId Default SampleID for a new measurement. ×
SampleInitial
Dilution
1:1 Default dilution ratio for a new measurement.
(1:2, 1:20, etc.).
×
SampleVolume 300 Default sample volume for a new measurement
(not adjustable).
×
SuggestNext
CupPosition
1 Increase in cup position after every measurement, for
example, when running new measurements in the
Measurement dialog box.
Possible values:
0=don’t increase, 1 = automatically increase.
Workarea Standard Default Workarea for a new measurement. ×
System Options and User Preferences