D
IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL 
8-88 Section 8-5 - Electronic Cage Components - Replacement Procedures
PRELIMINARY
8-5-2 Cabinet Cage Module Replacement Procedure
8-5-2-1 Tools
Phillips screwdriver and a 3mm Allen key.
8-5-2-2 Time Required
45 mins
8-5-2-3 Preparation
Shut down the Vivid S60/Vivid S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-5-2-4 Cabinet Cage Module Removal Procedure
1) Remove the following covers: left side, right side, DVD and Printer, front, right rear, left rear:
2) Remove the Front End Door Assembly, BEP, and Sub-Woofer Assembly:
3.) Disconnect all cables from the Internal BIF Panel located above the Cabinet Cage Module - see 
Figure 8-112.
Refer to Table 9-6 on page 9-5.
• Air Inlet (Left Side) Cover Removal Procedure on page 8 - 7
• Right Side Cover Removal Procedure on page 8 - 10
• DVD and Printer Cover Removal Procedure on page 8 - 13
• Front Cover Removal Procedure on page 8 - 16
• Right Rear Cover Removal Procedure on page 8 - 20
• Left Rear Cover Removal Procedure on page 8 - 23
• Front End Door Assembly Opening and Removal Procedure on page 8 - 83
• BEP Removal Procedure on page 8 - 94
• Sub-Woofer Assembly Removal Procedure on page 8 - 133
Figure 8-112   Disconnect Cables from Internal BIF Panel