8.3 Compensating Values to Match Reference Values (Load Compensation)
240
Setting the DC Bias
If the compensation frequency, compensation range, and compensation signal level are
not set, the DC bias setting cannot be set.
When DC is selected for the compensation frequency setting, the DC bias setting cannot
be set.
If you make a mistake during input:
press to cancel the input and start again.
DC Bias Setting
2. Select ON/OFF for the DC bias.
Sets the DC bias to OFF.
Sets the DC bias to ON.
3. Use or to enter the DC bias value.
Settable range: -5.00 V to 5.00 V (Normal mode)
-2.50 V to 2.50 V (low Z high accuracy
mode)
4. Press to close the setting screen.
Load Compensation Setting
1. Press .