8.3 Compensating Values to Match Reference Values (Load Compensation)
241
8
Chapter 8 Error Compensation
Setting of Parameter to Use for Reference Value
• If the compensation frequency, compensation range, and compensation signal level are
not set, the parameter to use for the reference value setting cannot be set.
• When DC is selected for the compensation frequency setting, DC resistance measure-
ment (Rdc) is selected automatically and the parameter to use for the reference value
setting cannot be set.
• If you change the parameter to use as the reference value, the settings of reference value
1 and reference value 2 are cleared.
Parameter Setting
See "1.3.7 Parameter Settings Screen" (p. 28)
2. Select the parameter mode of the reference value to be
set.
3. Press to close the setting screen.
Load Compensation Setting
1. Press .