EasyManua.ls Logo

iTero Element 5D Plus - Page 191

iTero Element 5D Plus
208 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Test Standard Class / Severity level
Test
results
Immunity from voltage dips, short
interruptions and voltage variations
IEC 61000-4-
11 /
EN 61000-4-11
On AC mode (240 V @ 50 Hz,
100 V @ 50 Hz): 0 % - 0.5 cycle
& 1 cycle;
70% - 25 cycles; 0% - 250
cycles;
On AC mode (220 V @ 60 Hz):
0 % - 0.5 cycle & 1 cycle;
70% - 30 cycles; 0% - 300
cycles
Complies
Emission (per ETSI EN 301 489-1, ETSI EN 301 489-17)
Conducted emissions on mains
terminals in freq. range 150 kHz - 30
MHz
ETSI EN 301
489-1;
ETSI EN 301
489-17 /
EN 55032
Group 1 Class B
230 VAC mains
Complies
Radiated emissions
in freq. range 30 - 6000 MHz
ETSI EN 301
489-1;
ETSI EN 301
489-17 /
EN 55032
Class B Complies
Harmonic current test ETSI EN 301
489-1;
ETSI EN 301
489-17 /
EN 61000-3-2
230 VAC mains Complies
Flicker tests ETSI EN 301
489-1 /
EN 61000-3-3
230 VAC mains Complies
Immunity (per ETSI EN 301 489-1, ETSI EN 301 489-17)
Immunity from Electrostatic
discharge (ESD)
EN 61000-4-2 4 kV contact discharge
8 kV air discharge
Complies
Immunity from radiated
electromagnetic fields
EN 61000-4-3 3.0 V/m,
80 MHz ¸ 6.0 GHz,
80% AM, 1 kHz
Complies
iTero Element® 5D Plus User manual
© 2020 Align Technology, Inc. All rights reserved. 173