Software Version C
- 112 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
Fig. 115 Zero point setting mode
Note!
The FACTORY zero point should be selected, if low turbidities are to be
measured exactly but no appropriate medium is available for zero point setting
in the process (user).
The factory zero point can be reactivated at any time.
The user's zero point is not deleted when changing from the factory zero point
and is maintained until the zero point is imported again by the user.
11. Save the setting. The SYSTEM SETTINGS menu appears.
SENSOR TF
FUNCTION TURB A/C
SENSOR TF16
MEDIUM FTU
ZERO POINT FACTORY
USER
REJECT P01 SAVE