Software Version C
- 171 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
9.12.2.2 Status feedback via the Remote Out 2 function
The Remote Out 2 function enhances the Remote Out 1 function to three
products. The relay outputs REL01, REL02, and REL03 are used for status
feedback. The relay outputs REL01 and REL02 at the clamps 10-11 and 12-13
are only activated when products P01, P02 or P03 are used.
The relay output REL03 at the clamps 14-15 is only activated in the measuring
mode.
When the CONTACTS NORMAL are set to OPEN, the relays switch as follows:
Relay 3: closed measuring mode
in MEASUREMENT DISPLAY
in DATA LOGGER DISPLAY
in SYSTEM INFORMATION DISPLAY
open measurement currently impossible because of
enabled PRODUCT CHANGE (3 sec.)
enabled PRODUCT CONFIGURATION
enabled SYSTEM SETTINGS
enabled MAINTENANCE
enabled HOLD
enabled PC TRANSFER
When the CONTACTS NORMAL are set to CLOSED, the relays switch as
follows:
Tab. 15 Relay status in Remote Out 2
Product P01 P02 P03 P04 to P08
Relay 1 closed open closed open
Relay 2 open closed closed open
Tab. 16 Relay status in Remote Out 2
Product P01 P02 P03 P04 to P08
Relay 1 open closed open closed
Relay 2 closed open open closed