9.14.4 Messages triggering system alarm
A system alarm is triggered when a system error can cause imprecise or wrong
measuring results. System alarm is also triggered when technically relevant
functions cannot be executed properly anymore.
System alarm is triggered by:
all lamp failures (1001 - 1011)
communication errors (1101 – 1105, 2011, 2012)
erroneous or unreadable hardware calibration data (1710 – 1754)
erroneous or unreadable system settings data (3201 – 3220)
erroneous or unreadable product configuration data,
linearizations and sensor adaptations (3230 – 3250).
Error messages regarding SIGNAL LOSS cannot trigger a system alarm,
since it is an admissible operational state in many applications!