Software Version C
- 146 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
9.11.7 Calibration mA-outputs
With this function, you can calibrate the mA-outputs anew. The calibration of
mA-outputs is carried out at 0 % and 80 % of the range.
When the mA-outputs are set to the basic setting of 4 - 20 mA (in the
mA OUTPUT menu item in SYSTEM SETTINGS see chapter 9.10.9, page 113),
the calibration range of 0 % and 80 % represents the values 4.00 and 16.80 mA.
When the mA-outputs are set to the the basic setting 0-20 mA, the calibration
range of 0 % and 80 % represents the values 0.00 and 16.00 mA.
Calibration with
factory settings
1. Select MAINTENANCE from the main menu and press [ENTER].
2. Select CALIBRATION mA OUTPUTS and press [ENTER].
Fig. 157 Menu maintenance, choosing calibration mA-outputs
3. Press [ENTER] in the CALIBRATION mA OUTPUTS menu. A list appears.
Select an output you want to calibrate. Press [ENTER].
Fig. 158 Calibration mA-outputs - choosing the mA-output
4. The cursor moves to the CALIBRATION menu item. Press [ENTER]. A list
appears. Select FACTORY SETTINGS and press [ENTER].
5. Save the setting.
MAINTENANCE
HOLD MANUALLY
ZERO POINT SETTING MANUALLY
SYSTEM DIAGNOSTICS
SENSOR ADAPTATION
CALIBRATION SCATTERED LIGHT
CALIBRATION mA INPUTS
CALIBRATION mA OUTPUTS
SERVICE
20:05:2006 P01 11:26:55
CALIBRATION mA OUTPUTS
mA OUTPUT : OUT 1
RANGE : 4 - 20 mA
CALIBRATION : NEW
SETPOINT ACTUAL
0 % : 4.00
80 % : 16.80
REJECT P01 SAVE