Software Version C
- 102 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
9.10.3 Data logger
With this function, you can set the measuring results storage interval and the
time resolution of the trend chart or reset the data logger. You find notes
regarding the data logger display in chapter 9.8, page 89.
1. Select SYSTEM SETTINGS from the main menu and press [ENTER].
2. Select DATA LOGGER and press [ENTER].
Fig. 101 Menu system settings, choosing data logger
3. Select INTERVAL and press [ENTER]. A list appears.
Fig. 102 Entering the storage interval
SYSTEM SETTINGS
DISPLAY
DATE / TIME
DATA LOGGER
DAMPING
LAMP VOLTAGE
SYSTEM ZERO POINT
PRODUCT CHANGE
SENSOR TF
20:05:2006 P01 11:26:55
SYSTEM SETTINGS
SELECT THE INTERVAL YOU WANT TO
STORE DATA
CURRENT SETTING IS: 01 SEC
INTERVAL: 01 SEC
05 SEC
...
60 MIN
_ _ _
DATA LOGGER
RESET:
REJECT P01 SAVE