Software Version C
- 158 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
Note!
• LEDs and relays are independent of each other when the relay outputs
are set to Remote Out in the SYSTEM SETTINGS menu (see
chapter 9.10.10, page 115).
• The switching status of a LED is determined by limit value, alarm type and
current measuring result.
• The update is not synchronous with the red status LED.
• The Relay + LED Monitor shows a snapshot, because measuring
operation is not carried out (see "Information on the measuring mode",
chapter 9.2.2, page 41).
Note!
• The relays 1 to 3 can be set for limit values or status feedback.
• The relays 1 to 3 can be defined normally open or normally closed.
• The system relay 4 provides two active contacts. It changes the switching
status as soon as voltage is applied to the converter. It opens, if a
system error occurs. It closes when the measuring mode is o.k.
9.11.8.2.6 Remote In monitor
This menu is described in chapter 9.12.1.1, page 164.