Software Version C
- 54 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
9.5.3 Define measuring results
With this function, you can define the measuring results, e. g. measuring range,
unit and figure format for the selected product. You can specify up to four
measuring results.
You can determine all configurations for the first measuring result of the
currently selected product with the function M01.
The mA-outputs 1 to 4 and the results 1 to 4 in the data logger are assigned to
the measuring results M01 to M04. Measuring result M01 is assigned to
mA-output 1, measuring result M02 is assigned to mA-output 2 etc.
The measuring result is calculated as follows:
1. Calculation of the selected measuring function for evaluating the detector
signals.
2. Multiplication with one or several sensor adaptations as far as they are
selected in the SENSOR ADAPTATION menu (see "Sensor adaptation",
chapter 9.11.4, page 140).
3. Multiplication with the mA input, as far as it is selected in the MEASURING
RESULTS menu (see "Measuring result - multiplication with mA-input",
chapter 9.5.3.3, page 63).
4. Multiplication with one of the 16 linearizations, as far as it is selected in the
MEASURING RESULTS menu (see "Measuring result - multiplication with
linearization", chapter 9.5.3.4, page 64).
5. Multiplication with the slope from the offset + slope set, as far as it is selected
in the MEASURING RESULTS menu (see "Measuring result - multiplication
with slope + offset", chapter 9.5.3.5, page 65).
6. Addition of the shift from the offset + slope set, as far as it is selected in the
MEASURING RESULTS menu (see "Measuring result - multiplication with
slope + offset", chapter 9.5.3.5, page 65).