EasyManua.ls Logo

Optek C4101 - Product Change, Setting the System Zero Point and the Hold Function in the Remote in 2 Function

Default Icon
208 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Software Version C
- 167 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
9.12.1.3 Product change, setting the system zero point and the hold function
in the Remote In 2 function
The Remote In 2 function treats the remote inputs R1-R4 differently in
dependence on R6. If R6 = LOW, all other inputs are treated as in Remote In 1.
If R6 = HIGH, the inputs R1 - R4 are used to zero the measuring results
M01 - M04 selectively, i. e. inputs A to D are only zeroed when occupied with
selected measuring results of the current product.
The signals should be allocated for at least one second to trigger a function.
Before triggering a function, a confirmation signal is always necessary
(R8 = HIGH = VALID).
"Remote In" inputs have the following functions:
•if R6 = low
• R1 + R2 + R3 = see table 12 = product selection P01 to P08
• R4 + R5 = low = not occupied
•if R6 = HIGH
• R1 + R2 + R3 + R4= see table 13 = measuring result selection M01
to M04
• R5 = low = not occupied
• R7, R8 and RR have the following functions:
• R7 = low = do not freeze system
• R7 = HIGH = freeze system as long as
R7 + R8 = HIGH
• R8 = HIGH = perform functions according to
R1 to R7
• RR = return of signals = returning signals

Table of Contents