9.12.1.3 Product change, setting the system zero point and the hold function
in the Remote In 2 function
The Remote In 2 function treats the remote inputs R1-R4 differently in
dependence on R6. If R6 = LOW, all other inputs are treated as in Remote In 1.
If R6 = HIGH, the inputs R1 - R4 are used to zero the measuring results
M01 - M04 selectively, i. e. inputs A to D are only zeroed when occupied with
selected measuring results of the current product.
The signals should be allocated for at least one second to trigger a function.
Before triggering a function, a confirmation signal is always necessary
(R8 = HIGH = VALID).
"Remote In" inputs have the following functions:
•if R6 = low
R1 + R2 + R3 = see table 12 = product selection P01 to P08
R4 + R5 = low = not occupied
•if R6 = HIGH
R1 + R2 + R3 + R4= see table 13 = measuring result selection M01
to M04
R5 = low = not occupied
R7, R8 and RR have the following functions:
R7 = low = do not freeze system
R7 = HIGH = freeze system as long as
R7 + R8 = HIGH
R8 = HIGH = perform functions according to
R1 to R7
RR = return of signals = returning signals