Software Version C
- 161 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
9.11.8.3 PC transfer
With this function, you can establish communication between the converter and
a PC via an RS-232 interface. With a PC and the CONTROL 4000 PC-
TRANSFER software, sold separately, you can perform the following functions:
Reading out the converter to a PC (read-out)
• Trend online for all 4 measuring results
• Data logger
• Parameter set
• Password protection and password allocation
•Formula set
• Calibration data scattered light
Reading in data from the PC to the converter (read-in)
• Parameter set
• Product configuration
• Linearization
• System settings
• Password protection and password allocation
•Formula set
• Calibration data scattered light
• Software update
System requirements
for PC transfer:
486 or Pentium
®
with Windows
®
- 95, - 98, - NT, - 2000, -XP or Vista.
Note!
• Copy the present data set from the converter to the PC before transferring
a new parameter set from the PC to the converter. This way, you can save
all old data sets!
• At the beginning of a transfer to the converter, the system automatically
replaces all existing parameters with default values.
• When the transfer is interrupted, the data sets are incomplete.
Parameters which have not been transferred are automatically replaced