EasyManua.ls Logo

Optek C4101 - PC Transfer

Default Icon
208 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Software Version C
- 161 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
9.11.8.3 PC transfer
With this function, you can establish communication between the converter and
a PC via an RS-232 interface. With a PC and the CONTROL 4000 PC-
TRANSFER software, sold separately, you can perform the following functions:
Reading out the converter to a PC (read-out)
• Trend online for all 4 measuring results
• Data logger
• Parameter set
• Password protection and password allocation
•Formula set
• Calibration data scattered light
Reading in data from the PC to the converter (read-in)
• Parameter set
• Product configuration
• Linearization
• System settings
• Password protection and password allocation
•Formula set
• Calibration data scattered light
• Software update
System requirements
for PC transfer:
486 or Pentium
®
with Windows
®
- 95, - 98, - NT, - 2000, -XP or Vista.
Note!
• Copy the present data set from the converter to the PC before transferring
a new parameter set from the PC to the converter. This way, you can save
all old data sets!
• At the beginning of a transfer to the converter, the system automatically
replaces all existing parameters with default values.
• When the transfer is interrupted, the data sets are incomplete.
Parameters which have not been transferred are automatically replaced
with default values. In this case, repeat the transfer.